PITOT TUBE BASIC INFORMATION


WHAT IS A PITOT TUBE AND WHAT ARE THE USES OF PITOT TUBE?



The pitot tube is used to measure the difference between the impact and static pressures in a fluid. It normally consists of two concentric tubes arranged parallel to the direction of flow; the impact pressure is measured on the open end of the inner tube.  

The end of the outer concentric tube is sealed and a series of orifices on the curved surface give an accurate indication of the static pressure. The position of these orifices must be carefully chosen because there are two disturbances which may cause an incorrect reading of the static pressure.

These are due to:

(1) the head of the instrument;
(2) the portion of the stem which is at right angles to the direction of flow of the fluid.

These two disturbances cause errors in opposite directions, and the static pressure should therefore be measured at the point where the effects are equal and opposite. If the head and stem are situated at a distance of 14 diameters from each other as on the standard instrument/41 the two disturbances are equal and opposite at a section 6 diameters from the head and 8 from the stem.

This is, therefore, the position at which the static pressure orifices should be located. If the distance between the head and the stern is too great, the instrument will be unwieldy; if it is too short, the magnitude of each of the disturbances will be relatively great, and a small error in the location of the static pressure orifices will appreciably affect the reading.

The two standard instruments are shown in Figure 6.13; the one with the rounded nose is preferred, since this is less subject to damage. For Reynolds numbers of 500-300,000, based on the external diameter of the pitot tube, an error of not more than 1 per cent is obtained with this instrument.

A Reynolds number of 500 with the standard 7.94 mm pitot tube corresponds to a water velocity of 0.070 m/s or an air velocity of 0.91 m/s. Sinusoidal fluctuations in the flowrate up to 20 per cent do not affect the accuracy by more than 1 per cent, and calibration of the instrument is not necessary.

A very small pressure difference is obtained for low rates of flow of gases, and the lower limit of velocity that can be measured is usually set by the minimum difference in pressure that can be measured. This limitation is serious, and various methods have been adopted for increasing the reading of the instrument although they involve the need for calibration.

Correct alignment of the instrument with respect to the direction of flow is important; this is attained when the differential reading is a maximum.

For the flow not to be appreciably disturbed, the diameter of the instrument must not exceed about one-fiftieth of the diameter of the pipe; the standard instrument (diameter 7.94 mm) should therefore not be used in pipes of less than 0.4 m diameter.

An accurate measurement of the impact pressure can be obtained using a tube of very small diameter with its open end at right angles to the direction of flow; hypodermic tubing is convenient for this purpose. The static pressure is measured using a single piezometer tube or a piezometer ring upstream at a distance equal approximately to the diameter of the pipe: measurement should be made at least 50 diameters from any bend or obstruction.

The pilot tube measures the velocity of only a filament of fluid, and hence it can be used for exploring the velocity distribution across the pipe section. If, however, it is desired to measure the total flow of fluid through the pipe, the velocity must be measured at various distances from the walls and the results integrated.

The total flowrate can be calculated from a single reading only if the velocity distribution across the section is already known. Although a single pitot tube measures the velocity at only one point in a pipe or duct, instruments such as the averaging pitot tube or Annubar, which employ multiple sampling points over the cross-section, provide information on the complete velocity profile which may then be integrated to give the volumetric flowrate. An instrument of this type has the advantage that it gives rise to a lower pressure drop than most other flow measuring devices, such as the orifice meter.

1 comment: